• Pharmacotherapy · Nov 2018

    Observational Study

    Accuracy of Point-of-Care Blood Glucose Level Measurements in Critically Ill Patients with Sepsis Receiving High-Dose Intravenous Vitamin C.

    • Kathryn E Smith, Caitlin S Brown, Bridget M Manning, Teresa May, Richard R Riker, Patricia A Lerwick, Timothy L Hayes, and Gilles L Fraser.
    • Department of Pharmacy, Maine Medical Center, Portland, Maine.
    • Pharmacotherapy. 2018 Nov 1; 38 (11): 1155-1161.

    Study ObjectiveHigh-dose intravenous vitamin C is a potential treatment option for patients with sepsis and may interfere with point-of-care (POC) blood glucose (BG) testing. This study aimed to determine if vitamin C dosing used for sepsis affected POC BG level results.DesignProspective observational pilot study.SettingIntensive care unit in a large academic tertiary care medical center.PatientsFive consecutive critically ill adults hospitalized between April 1 and June 1, 2017, who received two or more doses of intravenous vitamin C 1500 mg for the treatment of sepsis and had at least two paired POC BG levels and laboratory venous BG levels measured within 1 hour of each other during vitamin C therapy.Measurements And Main ResultsThe performance of POC BG level measurement was compared with the reference method of laboratory BG level measurement. The concordance to minimum accuracy criteria for BG meters set forth by the International Organization for Standardization (ISO) 15197:2013, the measurement of agreement between POC BG level and laboratory BG level using the Bland-Altman method, and the clinical accuracy through Parkes error grid analysis were assessed. A total of 16 paired POC and laboratory BG level measurements from the five patients were included. The accuracy of POC BG with laboratory BG level measurements during vitamin C administration according to ISO 15197:2013 criteria was 81.3%, which did not meet the minimum accuracy criteria of 95%. The Bland-Altman analysis showed a mean difference between POC and laboratory BG levels of 8.9 mg/dl, and the Parkes error grid analysis showed that the differences between POC and laboratory BG level measurements would not have resulted in a change in clinical action.ConclusionThe accuracy and agreement of POC and laboratory BG level measurements in critically ill patients receiving vitamin C were consistent with previously published reports in critically ill patients not receiving vitamin C and did not demonstrate clinically significant interference due to vitamin C dosing for sepsis.© 2018 Pharmacotherapy Publications, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…