ReviewMetrology part 2: Procedures for the validation of major measurement quality criteria and measuring instrument properties.
- Pierre Squara, ScheerenThomas W LTWLhttp://orcid.org/0000-0002-9184-4190Department of Anesthesiology, University of Groningen, University Medical Centre Groningen, Groningen, The Netherlands. t.w.l.scheeren@umcg.nl., Hollmann D Aya, Jan Bakker, Maurizio Cecconi, Sharon Einav, MalbrainManu L N GMLNGhttp://orcid.org/0000-0002-1816-5255Department of Intensive Care, University Hospital Brussels (UZB), Jette, Belgium.Faculty of Medicine and Pharmacy, Vrije Universiteit Brussel (VUB), Brussels, Belgium., Xavier Monnet, Daniel A Reuter, van der HorstIwan C CICChttp://orcid.org/0000-0003-3891-8522Department of Intensive Care, Maastricht University Medical Center+, Maastricht, The Netherlands., and Bernd Saugel.
- Department of Cardiology and ICU, Clinique Ambroise Paré, Neuilly-sur-Seine, France.
- J Clin Monit Comput. 2021 Feb 1; 35 (1): 27-37.
AbstractA measurement is always afflicted with some degree of uncertainty. A correct understanding of the different types of uncertainty, their naming, and their definition is of crucial importance for an appropriate use of the measuring instruments. However, in perioperative and intensive care medicine, the metrological requirements for measuring instruments are poorly defined and often used spuriously. The correct use of metrological terms is also of crucial importance in validation studies. The European Union published a new directive on medical devices, mentioning that in the case of devices with a measuring function, the notified body is involved in all aspects relating to the conformity of the device with the metrological requirements. It is therefore the task of scientific societies to establish the standards in their area of expertise. After adopting the same understandings and definitions (part 1), the different procedures for the validation of major quality criteria of measuring devices must be consensually established. In this metrologic review (part 2), we review the terms and definitions of validation, some basic processes leading to the display of an indication from a physiologic signal, and procedures for the validation of measuring instrument properties, with specific focus on perioperative and intensive care medicine including appropriate examples.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.