• Acta Anaesthesiol Scand · Jan 1976

    The flow requirement in a non-polluting Mapleson C circuit.

    • K N Christensen.
    • Acta Anaesthesiol Scand. 1976 Jan 1; 20 (4): 307-12.

    AbstractCapnography and measurement of the respiratory minute volume were carried out on normal, conscious subjects during breathing of pure O2 via a Mapleson C modification of a Magill attachment, with and without continuous gas evacuation. The measurements were performed in two series, with intermittent discharge and with continuous evacuation. Rebreathing was calculated in relation to the test subject's basal respiratory minute volume, measured at the commencement of the test. The measurements indicated CO2 accumulation at a flow below 200% of the respiratory minute volume. There was no significant difference between the systems with and without evacuation.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.