- Colleen M McFaul, Stefan Lombaard, Vivek Arora, William C Van Cleve, G Alec Rooke, and Jordan M Prutkin.
- From the Department of Anesthesiology and Pain Medicine, University of Ottawa Canada.
- A A Pract. 2020 Apr 1; 14 (6): e01178.
AbstractWe present the case of a patient with a subcutaneous implantable cardioverter-defibrillator (S-ICD) in situ. Device interrogation and reprogramming were unsuccessful due to a software mismatch between the device and programmer. The device manufacturer recommended magnet application to suspend antitachycardia therapy. Despite using this strategy, the S-ICD discharged multiple times. The S-ICD has unique perioperative considerations for the anesthesiologist. This case provides an example of the complexity of electrophysiologic devices in current use and the necessity of the anesthesia provider to stay up to date with evolving device management strategies.