Comparative Study
Influence of measurement principle on total hemoglobin value.
- Keisuke Hayashi, Takashi Hitosugi, Yoshifumi Kawakubo, Norihisa Kitamoto, and Takeshi Yokoyama.
- Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan.
- BMC Anesthesiol. 2020 Apr 7; 20 (1): 81.
BackgroundTotal hemoglobin (tHb) measurement is indispensable for determining the patient's condition (hemorrhagic vs. ischemic) and need for blood transfusion. Conductivity- and absorbance-based measurement methods are used for blood gas analysis of tHb. For conductivity-based measurement, tHb is calculated after converting blood conductivity into a hematocrit value, whereas absorbance measurement is based on light absorbance after red blood cell hemolysis. Due to changes in plasma electrolytes and hemolysis, there is a possibility that conductivity- and absorbance-based measurement methods may cause a difference in tHb.MethodsIn this study, test samples with controlled electrolyte changes and hemolysis were created by adding sodium chloride, distilled water or hemolytic blood to blood samples collected from healthy volunteers, and tHb values were compared between both methods.ResultsConductivity-based measurement revealed reduced tHb value (from 15.49 to 13.05 g/dl) following the addition of 10% sodium chloride, which was also reduced by the addition of hemolysate. Conversely, the addition of distilled water significantly increased tHb value than the expected value. In the absorbance method, there was no significant change in tHb value due to electrolyte change or hemolysis.ConclusionsWe have to recognize unexpected conductivity changes occur at all times when tHb is measured via conductivity- and absorbance-based measurement methods. The absorbance method should be used when measuring tHb in patients with expected blood conductivity changes. However, when using this method, the added contribution of hemoglobin from hemolytic erythrocytes lacking oxygen carrying capacity must be considered. We recognize that discrepancy can occur between conductivity- and absorbance-based measurement methods when tHb is measured.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.