• Nursing in critical care · Mar 2019

    Unintentional endotracheal tube cuff deflation during routine checks: a simulation study.

    • Vikesh Patel, Emily J Hodges, Maryanne Za Mariyaselvam, Catherine Peutherer, and Peter J Young.
    • University of Cambridge School of Clinical Medicine, Cambridge, UK.
    • Nurs Crit Care. 2019 Mar 1; 24 (2): 83-88.

    BackgroundEndotracheal tube cuff pressures should be maintained between 20 and 30 cm H2 O to prevent the aspiration of subglottic secretions past the cuff. Guidance recommends regular monitoring of the cuff pressure, performed using a handheld manometer/inflator. Poor technique can lead to transient deflation of the cuff, leading to the bolus aspiration of upper respiratory tract secretions, tracheal colonization and, ultimately, ventilator-associated pneumonia.Aims And ObjectivesTo determine whether intensive care staff transiently deflate the endotracheal tube cuff to below 20 cm H2 O during routine cuff pressure checks when using a handheld manometer/inflator device.DesignThis was an exploratory simulation study.MethodsA sample of medical (n = 10) and nursing staff (n = 10), capable of caring for a tracheally intubated patient, participated in the study on a single day. A mannequin was intubated with a standard oral endotracheal tube with the cuff pressure set at 50 cm H2 O. Participants were required to check and correct the cuff pressure to the appropriate level with a manometer. The lowest attained and the final target pressures were recorded.ResultsThree doctors were unfamiliar with the manometer and did not attempt measurement. During cuff pressure readjustment, 59% (10/17) of participants transiently deflated the cuff below 20 cm H2 O and then re-inflated to attain the final pressure. Of these participants, four deflated the cuff pressure to 0 cm H2 O before re-adjusting it back into range. Most participants, 88% (15/17), corrected the final cuff pressure to between 20 and 30 cm H2 O.ConclusionsPoor technique when using the manometer led to unintentional cuff deflation during routine checks. In clinical practice, this could increase the risk of pulmonary aspiration and ventilator-associated pneumonia. Further research into alternatives for handheld manometers, such as automated continuous cuff pressure monitors, is warranted.Relevance To Clinical PracticeCuff deflations can easily occur during routine cuff pressure checks. Staff should be aware of the implications of cuff deflations and seek to improve training with manometers.© 2018 British Association of Critical Care Nurses.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…