• BMC anesthesiology · Mar 2017

    Endotracheal tubes and fluid aspiration: an in vitro evaluation of new cuff technologies.

    • Maryanne Z Mariyaselvam, Lucy L Marsh, Sarah Bamford, Ann Smith, Matt P Wise, and David W Williams.
    • Oral and Biomedical Sciences, School of Dentistry, Cardiff University, Heath Park, Cardiff, Wales, CF14 4XY, UK. maryannemariyaselvam@gmail.com.
    • BMC Anesthesiol. 2017 Mar 4; 17 (1): 36.

    BackgroundAspiration of subglottic secretions past the endotracheal tube (ETT) cuff is a prerequisite for developing ventilator-associated pneumonia (VAP). Subglottic secretion drainage (SSD) ETTs reduce aspiration of subglottic secretions and have demonstrated lower VAP rates. We compared the performance of seven SSD ETTs against a non-SSD ETT in preventing aspiration below inflated cuffs.MethodsETTs were positioned vertically in 2 cm diameter cylinders. Four ml of a standard microbial suspension was added above inflated cuffs. After 1 h, aspiration was measured and ETTs demonstrating no leakage were subjected to rotational movement and evaluation over 24 h. Collected aspirated fluid was used to inoculate agar media and incubated aerobically at 37 °C for 24 h. The aspiration rate, volume and number of microorganisms that leaked past the cuff was measured. Experiments were repeated (×10) for each type of ETT, with new ETTs used for each repeat. Best performing ETTs were then tested in five different cylinder diameters (1.6, 1.8, 2.0, 2.2 and 2.4 cm). Experiments were repeated as above using sterile water. Volume and time taken for aspiration past the cuff was measured. Experiments were repeated (×10) for each type of ETT. Results were analysed using non-parametric tests for repeated measures.ResultsThe PneuX ETT prevented aspiration past the cuff in all experiments. All other ETTs allowed aspiration, with considerable variability in performance. The PneuX ETT was statistically superior in reducing aspiration compared to the SealGuard (p < 0.009), KimVent (p < 0.002), TaperGuard (p < 0.004), Lanz (p < 0.001), ISIS (p < 0.001), SACETT (p < 0.001) and Soft Seal (p < 0.001) ETTs. Of the 4 ETTs tested in differing cylinder sizes, the PneuX significantly reduced aspiration across the range of diameters compared to the SealGuard (p < 0.0001), TaperGuard (p < 0.0001) and KimVent (p < 0.0001) ETTs.ConclusionsETTs showed substantial variation in fluid aspiration, relating to cuff material and design. Variability in performance was likely due to the random manner in which involutional folds form in the inflated ETT cuff. The PneuX ETT was the only ETT able to consistently prevent aspiration past the cuff in all experiments.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.