Randomized Controlled Trial
Pedicle screw insertion with patient-specific 3D-printed guides based on low-dose CT scan is more accurate than free-hand technique in spine deformity patients: a prospective, randomized clinical trial.
- Riccardo Cecchinato, Pedro Berjano, Alberto Zerbi, Marco Damilano, Andrea Redaelli, and Claudio Lamartina.
- GSpine4, IRCCS Istituto Ortopedico Galeazzi, Milan, Italy. dott.cecchinato@gmail.com.
- Eur Spine J. 2019 Jul 1; 28 (7): 1712-1723.
BackgroundScrew misplacement incidence can be as high as 15-30% in spine deformity surgery, with possible devastating consequences. Some technical solutions to prevent misplacement require expensive devices. MySpineTM comprises a low-dose CT scan of the patient's spine to build a virtual model of the spine to plan the screw trajectories and a 3D-printed patient-specific guide system to prepare the screw trajectories and to implant the screws in the vertebrae in order to increase reproducibility and safety of the implants. The aim of this open-label, single-center, prospective randomized clinical trial with independent evaluation of outcomes was to compare the accuracy of free-hand insertion of pedicle screws to MySpineTM 3D-printed patient-specific guides.MethodsTwenty-nine patients undergoing surgical correction for spinal deformity were randomized to Group A (pedicle screws implantation with MySpineTM) or Group B (free-hand implantation). Group A received 297 pedicle screws, and Group B 243 screws. Forty-three screws in Group A crossed over to free-hand implantation. Screw position was graded according to Gertzbein in grades 0, A, B or C, with grades 0 or A considered as "safe area." Total fluoroscopy dose and time were compared in six patients of each group.ResultsComparing the two study groups, we observed a statistically significant difference between the two groups (p < 0.05), with 96.1% of screws in the "safe area" in Group A versus a 82.9% in Group B. Group-A patients had a mean effective dose of 0.23 mSv compared to 0.82 mSv in Group B. Patient-specific, 3D-printed pedicle screw guides increase safety in a wide spectrum of deformity conditions. In addition, the total radiation dose is reduced, even considering the need of a low-dose preoperative CT for surgical planning.Level Of EvidenceI. These slides can be retrieved under Electronic Supplementary Material.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.