Gastrointestinal endoscopy
-
Gastrointest. Endosc. · Nov 2019
Independent root-cause analysis of contributing factors, including dismantling of 2 duodenoscopes, to investigate an outbreak of multidrug-resistant Klebsiella pneumoniae.
Worldwide, an increasing number of duodenoscope-associated outbreaks are reported. The high prevalence rate of contaminated duodenoscopes puts patients undergoing ERCP at risk of exogenous transmission of microorganisms. The contributing factors of the duodenoscope design to contamination are not well understood. This article reports on the investigation after the outbreak of a multidrug-resistant Klebsiella pneumoniae (MRKP) related to 2 Olympus TJF-Q180V duodenoscopes. ⋯ Outbreaks are associated with a combination of factors, including duodenoscope design issues, repair issues, improper cleaning, and systemic monitoring of contamination. To eliminate future duodenoscope-associated infections, a multipronged approach is required, including clear communication by all parties involved, a reliable servicing market, stringent surveillance measures, and eventually new duodenoscope designs and reprocessing procedures with a larger margin of safety.