Quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPECT: MRI validation of the EXINI heart software package.
The aim of the study was to validate the accuracy of the EXINI heart software (EXINI) package in assessing left ventricular end-diastolic/systolic volumes (EDV, ESV) and ejection fraction (LVEF) from gated (99m)Tc-MIBI single-photon emission tomography (SPECT). Cardiac magnetic resonance imaging (cMRI) was used as reference. Furthermore, effects of perfusion defects and image quality in SPECT on correlation between gated SPECT and magnetic resonance imaging were investigated. ⋯ End-diastolic volume, ESV and LVEF calculated from gated SPECT using EXINI agree with cMRI over a wide range of values. Correlation between both the methods was good for EDV and ESV, and acceptable for LVEF. No relevant influence of image quality or SRS on the accuracy of EXINI results was found.