Multicenter StudyThe accuracy of quantitative parameters in (99m) Tc-MAG3 dynamic renography: a national audit based on virtual image data.
Assessment of image analysis methods and computer software used in (99m) Tc-MAG3 dynamic renography is important to ensure reliable study results and ultimately the best possible care for patients. In this work, we present a national multicentre study of the quantification accuracy in (99m) Tc-MAG3 renography, utilizing virtual dynamic scintigraphic data obtained by Monte Carlo-simulated scintillation camera imaging of digital phantoms with time-varying activity distributions. Three digital phantom studies were distributed to the participating departments, and quantitative evaluation was performed with standard clinical software according to local routines. ⋯ The Tmax estimates showed a similar trend, but the difference was only significant for the right kidney. There was a significant bias in the measured DRF for all phantoms caused by different positions of the left and right kidney in the anterior-posterior direction. In conclusion, this study shows that virtual scintigraphic studies are applicable for quality assurance and that there is a considerable uncertainty associated with standard quantitative parameters in dynamic (99m) Tc-MAG3 renography, especially for patients with impaired renal function.