Comparative StudyMeasuring end-expiratory lung volume and pulmonary mechanics to detect early lung function impairment in rabbits.
We investigated whether end-expiratory lung volume (EELV) or lung mechanical parameters are more sensitive for the detection of a compromised gas exchange during bronchoconstriction and after surfactant depletion. EELV was determined via SF(6) multiple breath wash-outs in mechanically ventilated rabbits while a positive end-expiratory pressure (PEEP) of 1, 3 or 7 cm H(2)O was maintained. Airway resistance (R(aw)) and parenchymal elastance (H) were estimated from the pulmonary input impedance measured at each PEEP level by means of forced oscillations. ⋯ MCh induced marked elevations in R(aw), with no significant change in EELV or H at any PEEP. After lavage, the severity of hypoxia was reflected systematically in significant decreases in EELV at all PEEP levels (-42+/-13%, -26+/-4%, and -18+/-5% at 1, 3 and 7 cm H(2)O, respectively), whereas compromised gas exchange was not associated with consistent changes in the mechanical parameters at a PEEP of 7 cm H(2)O (20+/-9% and 14+/-9% in R(aw) and H, respectively; p=0.2). We conclude that R(aw) is the only sensitive indicator for the detection of a compromised lung function during MCh infusions, whereas the estimation of EELV is necessary to follow the progression of a lung injury when a high PEEP level is applied.