Respiratory physiology & neurobiology
-
Respir Physiol Neurobiol · Jul 2012
Clinical TrialThe measurement of carboxyhemoglobin and methemoglobin using a non-invasive pulse CO-oximeter.
The pulse CO-oximeter (Rad-57 Masimo Corporation, Irvine, CA) allows non-invasive and instantaneous measurement of carboxyhemoglobin (COHb) and methemoglobin (MetHb) percentage level using a finger probe. However, the accuracy and reliability of the Rad-57 against the gold standard of venous or arterial blood samples have not been clearly established. Thus, the objective of this trial is to evaluate the accuracy and precision of the Rad-57 pulse CO-oximeter by comparing it with venous sampling on the same subjects. ⋯ The ability to detect a COHb ≥ 10% occurred in 54% of the samples in which COHb was ≥ 10-14%. In conclusion, the Rad-57 provides a reading that is between -6% and +4% of the true COHb value for 95% of all samples. The Rad-57 seems to be a good substitute as a first screening test of COHb when the pulse CO-oximeter reads <15%.