Randomized Controlled Trial Clinical Trial[Continuous positive airway pressure (CPAP)--does the equipment live up to its name?].
The purpose of this study was to evaluate the in- and expiratory pressures achieved by the use of different continuous positive airway pressure (CPAP) delivery systems. The study was performed on ten healthy adults breathing spontaneously with: 1) "simple CPAP", consisting of a fresh gas flow around 40 l/min and reservoir balloons of 2, 6 and 10 l and expiration against a water column, 2) a high-flow system CR60 CPAP, designed for home treatment and 3) another high-flow system Down's CPAP. Both high-flow systems used a venturi device (Downs Flow Generator, Vital Signs) giving flows around 90 and 170 l/min in our set-up, and the same PEEP valves (Vital signs) were used. ⋯ The high-flow systems were not found to provide better CPAP, however, CR60 CPAP was found also to be optimal with 5 and 10 cm H2O. Down's CPAP provided 3-5 cm higher expiratory pressure than the intended CPAP level. We recommend measurement of the pressure during the CPAP treatment to ensure that the intended CPAP is achieved.
-
Ugeskrift for laeger · May 1992
[Guanethidine blockade for palliative treatment in reflex dystrophy].
Guanethidine blockade was employed in 22 patients with causalgia, hyperpathia and reflex dystrophy in an upper limb. Guanethidine was administered by means of regional intravenous technique in which 20 mg guanethidine was mixed with 400 IU of heparin and 70 mg lignocaine. ⋯ In 68% of the patients, beneficial results occurred as they were relieved of pain or improved considerably. Employment of lignocaine in the injection solution made the blockade less painful.