Applied health economics and health policy
-
Appl Health Econ Health Policy · Sep 2012
ReviewPleurX peritoneal catheter drainage system for vacuum-assisted drainage of treatment-resistant, recurrent malignant ascites: a NICE Medical Technology Guidance.
The PleurX peritoneal drainage catheter for drainage of malignant ascites in a community setting has been evaluated by the NICE Medical Technologies Evaluation Programme. This article outlines the evidence included in the Sponsor's submission, the independent critique by the External Assessment Centre (EAC) and the recommendations made by the Medical Technologies Advisory Committee (MTAC). In accordance with the scope issued by NICE, the intervention technology was the indwelling PleurX peritoneal catheter drainage system, the comparator was large-volume paracentesis (LVP; inpatient or outpatient) and the population was patients with treatment-resistant, recurrent malignant ascites. ⋯ There was uncertainty surrounding the number of patients for whom insertion of a PleurX drain would be appropriate as well as the point in the care pathway at which such treatment should be administered. MTAC supported the case for adoption and considered that the available evidence showed PleurX was clinically effective, has low complication rates, can improve quality of life and is less costly than inpatient LVP. In Medical Technology Guidance 9 (MTG9), NICE recommended that PleurX peritoneal catheter drainage system be considered for use in patients with treatment-resistant, recurrent malignant ascites.