A novel approach using time-frequency analysis of pulse-oximeter data to detect progressive hypovolemia in spontaneously breathing healthy subjects.
Accurate and early detection of blood volume loss would greatly improve intraoperative and trauma care. This study has attempted to determine early diagnostic and quantitative markers for blood volume loss by analyzing photoplethysmogram (PPG) data from ear, finger and forehead sites with our high-resolution time-frequency spectral (TFS) technique in spontaneously breathing healthy subjects (n = 11) subjected to lower body negative pressure (LBNP). The instantaneous amplitude modulations present in heart rate (AM HR) and breathing rate (AMBR) band frequencies of PPG signals were calculated from the high-resolution TFS. ⋯ The mean percent increase in AMBR values at 100% LBNP tolerance was 99.4% and 19.6% for ear and finger sites, respectively; AMBR values were not attainable for forehead PPG signal. Even without baseline AMHR values, our results suggest that hypovolemia detection is possible with specificity and sensitivity greater than 90% for the ear and forehead locations when LBNP tolerance is 100%. Therefore, the TFS analysis of noninvasive PPG waveforms is promising for early diagnosis and quantification of hypovolemia at levels not identified by vital signs in spontaneously breathing subjects.