Journal of thrombosis and haemostasis : JTH
-
J. Thromb. Haemost. · Aug 2021
Evaluation of laboratory assays for anti-platelet factor 4 antibodies after ChAdOx1 nCOV-19 vaccination.
Vaccine-induced immune thrombocytopenia and thrombosis (VITT) following ChAdOx1 nCOV-19 vaccine has been described, associated with unusual site thrombosis, thrombocytopenia, raised D-dimer, and high-titer immunoglobulin-G (IgG) class anti-platelet factor 4 (PF4) antibodies. Enzyme-linked immunosorbent assays (ELISA) have been shown to detect anti-PF4 in patients with VITT, but chemiluminescence assays do not reliably detect them. ELISA assays are not widely available in diagnostic laboratories, and, globally, very few laboratories perform platelet activation assays. ⋯ We demonstrated that the HemosIL AcuStar HIT-IgG, HemosIL HIT-Ab, Diamed PaGIA gel, and STic Expert assays have poor sensitivity for VITT in comparison to ELISA. Where these assays are used for heparin-induced thrombocytopenia (HIT) diagnosis, laboratories should ensure that requests for suspected VITT are clearly identified so that an ELISA is performed. No superiority of IgG-ELISAs over polyspecific ELISAs in sensitivity to VITT could be demonstrated. No single ELISA method detected all possible/probable VITT cases; if a single ELISA test is negative, a second ELISA or a platelet activation assay should be considered where there is strong clinical suspicion.