Kinetics of isoflurane and sevoflurane in a unidirectional displacement flow and the relevance to anesthetic gas exposure by operating room personnel.
International guidelines recommend the use of ventilation systems in operating rooms to reduce the concentration of potentially hazardous substances such as anesthetic gases. The exhaust air grilles of these systems are typically located in the lower corners of the operating room and pick up two-thirds of the air volume, whereas the final third is taken from near the ceiling, which guarantees an optimal perfusion of the operating room with a sterile filtered air supply. However, this setup is also employed because anesthetic gases have a higher molecular weight than the components of air and should pool on the floor if movement is kept to a minimum and if a ventilation system with a unidirectional displacement flow is employed. ⋯ Interestingly, neither isoflurane nor sevoflurane was detected at floor level, but a continuous cone-like spreading was observed due to gravity. In contrast, carbon dioxide accumulated at floor level in the form of a gas cloud. Thus, floor level exhaust ventilation systems are likely unsuitable for the collection and removal of anesthetic gases from operating rooms.