Heart rhythm : the official journal of the Heart Rhythm Society
-
Predictable progression to battery depletion is necessary for device management in patients with pacemakers or implantable cardioverter-defibrillators, particularly in patients who either are pacemaker dependent or have required implantable cardioverter-defibrillator therapies. ⋯ The incidence of abrupt battery depletion was 4.2% in patients implanted with a St Jude Atlas CRT-D device. No common mechanism has been identified for this failure. Close monitoring of battery voltage and timely generator replacement are required in patients with these devices.
-
Cardiac sodium channel β-subunit mutations have been associated with several inherited cardiac arrhythmia syndromes. ⋯ Our results suggest that R214Q variation in SCN1Bb is a functional polymorphism that may serve as a modifier of the substrate responsible for BrS or SIDS phenotypes via a combined loss of function of sodium channel current and gain of function of transient outward potassium current.
-
The mechanism(s) underlying the maintenance of atrial fibrillation (AF) during the first few hours after AF was initiated remains poorly understood. ⋯ RAP not only induces atrial electrical remodeling but also promotes autonomic remodeling. These 2 remodeling processes may form a vicious cycle and each may perpetuate the other. These findings may help to explain how AF maintains itself in its very early stage. LL-SVCS both reversed remodeling processes and can potentially break the vicious cycle of "AF begets AF" in the first few hours of AF.