Polyethylene wear after total hip arthroplasty: the effect of a modular femoral head with an extended flange-reinforced neck.
The use of modular components in total hip arthroplasty has been thought to contribute to accelerated polyethylene wear. Specifically, a modular femoral head with a flange extension and a longer neck may cause increased wear. The purpose of the current study was to evaluate the effect of a flange extension on polyethylene wear. ⋯ Multivariate analysis showed that the presence of a flange extension was associated with increased polyethylene wear to a greater degree (F = 2.86) than were all other variables that were measured, including a younger age (F = 1.72), a more vertical angle of the acetabular component (F = 0.49), a heavier weight (F = 0.14), male gender (F = 0.11), and a smaller initial thickness of the polyethylene (F = 0.02). These data support an association between the presence of a modular femoral head with a flange extension and an accelerated rate of polyethylene wear. The presumed mechanism is an increase in peripheral, or so-called rim, impingement of the flange-reinforced neck on the acetabulum due to a decrease in the ratio between the diameters of the femoral head and neck.