To determine the oxygen outflow delivered by seven different models of manually operated self-inflating resuscitation bags (with and without an oxygen reservoir connected), which were tested using different oxygen supply rates without manipulating the bag, by simulating their use in patients breathing spontaneously. ⋯ Of the resuscitation bags studied, those that allow the connection of an oxygen reservoir must have this reservoir connected to the bag when used as a source of oxygen in nonintubated spontaneously breathing patients. All of the models studied should receive oxygen at a rate > 15 L/min. It is not safe to use manually operated self-inflating resuscitation bags for this purpose without knowing their characteristics.