Journal of neurosurgery. Spine
-
Review Meta Analysis
Biomechanical analysis of screw constructs for atlantoaxial fixation in cadavers: a systematic review and meta-analysis.
The unique and complex biomechanics of the atlantoaxial junction make the treatment of C1-2 instability a challenge. Several screw-based constructs have been developed for atlantoaxial fixation. The biomechanical properties of these constructs have been assessed in numerous cadaver studies. The purpose of this study was to systematically review the literature on the biomechanical stability achieved using various C1-2 screw constructs and to perform a meta-analysis of the available data. ⋯ Meta-analysis of the existing literature showed that all constructs provided significant stabilization in all axes of rotation, except for the C1LM-C2TL construct in lateral bending. There were significant differences in stabilization achieved in each axis of motion by the various screw constructs. These results underline the various strengths and weaknesses in biomechanical stabilization of different screw constructs. There was significant heterogeneity in the data reported across the studies. Standardized spinal motion segment configuration and injury models may provide more consistent and reliable results.