Journal of biomechanics
-
Journal of biomechanics · Sep 1992
Comparative StudyBiomechanical consequences of callus development in Hoffmann, Wagner, Orthofix and Ilizarov external fixators.
A theoretical analysis by a finite elements model (FEM) of some external fixators (Hoffmann, Wagner, Orthofix and Ilizarov) was carried out. This study considered a logarithmic progress of callus elastic characteristics. A standard configuration of each fixator was defined where design and application characteristics were modified. ⋯ A minimal contribution from an external fixator to the total rigidity of the bone-callus-fixator system was assessed when a callus showing minimum elastic characteristics had just been established. Insufficient rigidity from the fixation devices to assure an adequate immobilization during the early stages of fracture healing was verified. However, regardless of the external fixator, callus development was the overriding element for the rigidity of the fixator-bone system.