Bone cement augmentation of femoral nail head elements increases their cut-out resistance in poor bone quality- A biomechanical study.
The aim of this study was to analyze biomechanically the impact of bone cement augmentation on the fixation strength and cut-out resistance of Proximal Femoral Nail Antirotation (PFNA) and Trochanteric Fixation Nail Advanced (TFNA) head elements within the femoral head in a human cadaveric model with poor bone quality. Methodology: Fifteen pairs of fresh-frozen human cadaveric femoral heads were randomized to three sets of five pairs each for center-center implantation of either TFNA blade, TFNA screw, or PFNA blade. By splitting the specimens of each pair for treatment with or without bone cement augmentation, six study groups were created. ⋯ Augmented TFNA blades resulted in highest numbers of cycles to failure and loads at failure (30492; 4049 N) followed by augmented PFNA blades (30033; 4003 N) and augmented TFNA screws (19307; 2930 N), p = 0.11. Augmented TFNA screws showed similar numbers of cycles to failure and loads at failure compared to both non-augmented TFNA and PFNA blades, P = 0.98. From a biomechanical perspective, bone cement augmentation significantly increases the cut-out resistance of instrumented TFNA and PFNA head elements and is a valid supplementary treatment option to these nailing procedures in poor bone quality.