Measurement of the cardiac output in small animals by thermodilution.
Cardiac output (CO) measurements based on indicator dilution, microspheres, thermodilution and ultrasonic sensors are not suitable for small animals, because of limited blood volume, high heart rates and small caliber vessels that do not allow probe placement within the heart. We developed a modified thermodilution method to measure CO in awake animals weighing less than 100 g. Under anesthesia, the animal is instrumented with a jugular vein catheter placed proximal to the subclavian vein and a temperature probe in the carotid artery with the thermocouple positioned at the aortic arch. ⋯ Cardiac index (CI=CO/weight) was 197.0 +/- 18.8 (ml/min)/kg. Repeated measurements were made. This technique allows correlating systemic flow changes to be correlated to those measured in the microcirculation of window chamber preparations.