International journal of laboratory hematology
-
Comparative Study
Comparability of the results of PT-INR with local MNPT and APTTR with MNAPTT on different coagulation analyzers in China.
The prothrombin time (PT), International normalized ratio (INR) and activated partial thromboplastin time (APTT) are the most used coagulation tests in China, where more than one type of automated coagulation analyzer is often used in the clinical laboratory. The PT-INR results of 109 samples were compared with local mean normal PT (MNPT) and APTT ratio (APTTR) with mean normal APTT (MNAPTT) on two different coagulation analyzers in the same laboratory. ⋯ Meanwhile, there was no significant difference (P = 0.865) for APTTR with MNAPTT compared with APTT (P = 0.002) between the ACL Futura and CA 510. In conclusion, these analyzers showed very poor agreement for both the PT and APTT, but the calculation of ratios significantly improved agreement.