A device for temporary external DDD pacing was developed using a modified permanent pulse generator, and temporary atrial and ventricular electrode catheters were inserted pervenously. The atrial lead was a J type and the ventricular lead had a remote anode in the superior vena cava. With a special clamp and appropriate connectors the permanent DDD pulse generator provided unipolar external DDD pacing. ⋯ Eight patients benefited from temporary DDD pacing and 7 had permanent DDD pacemakers implanted. In 1 patient, atrial fibrillation developed, 2 patients had brief episodes of ventricular lead displacement and another required reprogramming because of loss of ventricular capture. Results suggest that the device is applicable for a clinical trial of DDD pacing before possible permanent implantation when attempting to improve cardiac output or control arrhythmias.