The American journal of cardiology
-
One hundred thirty-four patients with prosthetic or bioprosthetic heart valves were investigated with Doppler echocardiography to determine normal values for commonly used prosthetic valves and to test the specificity of abnormal Doppler findings. In 70 patients the aortic valves had been replaced and in 64 the mitral valves had been replaced. Gradients across prostheses in the aortic position were calculated from maximal velocity. ⋯ Mitral regurgitation was found in 11% of Björk-Shiley valves, 19% of porcine valves and 30% of Starr-Edwards valves. Repeat studies at 2 weeks to 14 months revealed no difference in 8 aortic and 14 mitral prostheses. Seven aortic and 4 mitral valves functioned abnormally as determined by Doppler, and the abnormal function was confirmed in each at surgery or at cardiac catheterization.(ABSTRACT TRUNCATED AT 250 WORDS)