Acute changes in pacing threshold and R- or P-wave amplitude during permanent pacemaker implantation.
This study examines the changes in pacing threshold and R- or P-wave amplitude during the first 30 minutes after implantation of tined and screw-in leads. The leads examined were those of 1 manufacturer (Medtronic) and consisted of 3 ventricular pacing leads (model numbers 6957 unipolar screw-in [11 patients], 6961 unipolar tined [12 patients] and 6962 bipolar tined [7 patients]) and 1 atrial lead (model number 6957J unipolar screw-in [10 patients]). After optimal lead position was obtained fluoroscopically in the right ventricular apex or right atrium, the pacing threshold and R- or P-wave amplitudes were measured at 5-minute intervals for 30 minutes. ⋯ There was a significant acute increase in R-wave size with the ventricular screw-in lead that peaked 20 minutes after lead implantation (11.9 +/- 3.0 to 14.7 +/- 4.1 mV; p less than 0.001). The atrial screw-in lead behaved in a manner identical to its counterpart in the ventricle. In conclusion, there are acute changes in the pacing threshold and R- or P-wave amplitude obtained with tined and screw-in pacing leads.(ABSTRACT TRUNCATED AT 250 WORDS)