Drug discoveries & therapeutics
-
Observational Study
Exploring the causes of peripheral intravenous catheter failure based on shape of catheters removed from various insertion sites.
The risk of peripheral intravenous catheter failure varies according to the insertion site. This study examined catheter shape just after removal to evaluate the causes of catheter failure according to site. This study was a secondary analysis of previous study data. ⋯ Catheter curvature was related to catheter failure (14.8% of failed catheters had curvature; p = 0.035) and occlusion (35.3% of occluded catheters had curvature; p = 0.008) in upper arm and forearm placements. The median distance from the elbow to the insertion site was shorter for failed catheters than for surviving catheters. To prevent catheter failure, especially occlusion resulting from catheter curvature, a catheter should be inserted at an appropriate insertion site far from the antecubital fossa.