Biomedizinische Technik. Biomedical engineering
-
Randomized Controlled Trial Comparative Study Clinical Trial
Auto-CPAP therapy based on the forced oscillation technique.
Autoadjusting CPAP devices (APAP) are designed to continuously adjust the positive pressure to the required levels, and thus increase treatment quality and patient compliance. The results of APAP treatment strongly depend on the control mechanism of the respective APAP device. In agreement with other working groups, we have recently shown that the forced oscillation technique (FOT) is capable of detecting incipient upper airway obstruction prior to physiological reactions such as the onset of increasing esophageal pressure swings or microarousals. ⋯ The mean treatment pressure during APAP was significantly lower than during CPAP treatment (6.0 +/- 2.0 vs. 9.0 +/- 1.8 mbar; p < 0.001). There were no significant differences between APAP and CPAP treatment in any parameter of efficacy or acceptance. APAP treatment with this device controlled exclusively by FOT is well accepted by the patients and permits an adequate treatment of OSAS without the need for invidiual CPAP titration.