Biomedizinische Technik. Biomedical engineering
-
In this paper, common time- and frequency-domain variability indexes obtained by pulse rate variability (PRV) series extracted from video-photoplethysmographic signal (vPPG) were compared with heart rate variability (HRV) parameters calculated from synchronized ECG signals. The dual focus of this study was to analyze the effect of different video acquisition frame-rates starting from 60 frames-per-second (fps) down to 7.5 fps and different video compression techniques using both lossless and lossy codecs on PRV parameters estimation. Video recordings were acquired through an off-the-shelf GigE Sony XCG-C30C camera on 60 young, healthy subjects (age 23±4 years) in the supine position. ⋯ Concerning the video compression, the results showed that compression techniques are suitable for the storage of vPPG recordings, although lossless or intra-frame compression are to be preferred over inter-frame compression methods. FFV1 performances are very close to the uncompressed (UNC) version with less than 45% disk size. H.264 showed a degradation of the PRV estimation directly correlated with the increase of the compression ratio.