The authors constructed a D-shaped tracheal model with an elastic posterior wall, thus simulating normal tracheal anatomy more closely than previous models. The performance of 9-10 tracheal tube cuffs, of 2-3 different tube sizes (7.0-10.0 mm, ID), from six different manufacturers were tested in the model. Cuff residual volumes ranged from 1.78 to 27.35 ml. ⋯ There was no relationship between cuff brands in the wall pressure required to effect a seal in the model. The authors conclude that intratracheal tubes should have cuffs with large residual volumes. This would permit some latitude in tube size selection while ensuring that a seal could be achieved before the cuff is inflated to its residual volume.
-
The incidence of neurologic complications arising from anticoagulant therapy, following epidural and subarachnoid catheterization in 3,164 and 847 patients, respectively, was determined. Twenty patients experienced minor neurologic complications or low back pain which was self-limiting and resolved with time. There was no incidence of peridural hematoma leading to spinal cord compression. This investigation shows that the occurrence of symptomatic hematomas following anticoagulation in patients with epidural or subarachnoid catheters is a very rare complication, assuming proper patient selection, an atraumatic technique, and appropriate monitoring of anticoagulant activity.