Plastic and reconstructive surgery
-
Plast. Reconstr. Surg. · Apr 1995
Case ReportsSupplemental oxygen: ensuring its safe delivery during facial surgery.
Electrosurgical coagulation in the presence of blow-by oxygen is a potential source of fire in facial surgery. A case report of a patient sustaining partial-thickness facial burns secondary to such a flash fire is presented. A fiberglass facial model is then used to study the variables involved in providing supplemental oxygen when an electrosurgical unit is employed. ⋯ The properly placed nasal cannula did not ignite at any combination of oxygen flow, coagulation current level, or distance from the oxygen source. Facial cutaneous surgery in patients provided supplemental oxygen should be practiced with caution when an electrosurgical unit is used for coagulation. The oxygen delivery systems adapted for use are hazardous and should not be used until their safety has been demonstrated.