Twenty-seven patients were examined who were undergoing cardiopulmonary bypass (CPB) surgery with either a bubble oxygenator or a capillary membrane oxygenator. The latter incorporated an arterial filter and bubble trap. A noninvasive Doppler ultrasound technique is described for monitoring irregularities in the Doppler flow signals attributable to gaseous microemboli detected in the middle cerebral artery during CPB. ⋯ However, all 10 patients with a membrane oxygenator had an MEI of 0. Varying the gas flow rates in 3 patients with bubble oxygenators showed a change in MEI from 4 +/- 4 (SD) at a flow rate of 2 L/min to 17 +/- 9 at a flow rate of 5 L/min. This observation supports the assumption that the MEI is providing quantitative information regarding the presence of gaseous emboli in the middle cerebral artery.