Health physics
-
Comparative Study Clinical Trial
Evaluation of patient and staff doses during various CT fluoroscopy guided interventions.
As CT scanners are more routinely used as a guidance tool for various types of interventional radiological procedures, concern has grown for high patient and staff doses. CT fluoroscopy provides the physician immediate feedback and can be a valuable tool to dynamically assist various types of percutaneous interventions. However, the fixed position of the scanning plane in combination with high exposure factors may lead to high cumulative patient skin doses that can reach deterministic threshold limits. ⋯ CT fluoroscopy doses are markedly higher than classic CT-scan doses and are comparable to doses from other interventional radiological procedures. They consequently require adequate radiation protection management. An important potential for dose reduction exists by limiting the fluoroscopic screening time and by reducing the tube current (mA) to a level sufficient to provide adequate image quality.