HNO
-
Laser surgery can be performed using either endotracheal intubation, apnoea or jet ventilation. For operations performed under endotracheal intubation the same technical rules apply as for any other type of anaesthesia. To facilitate intubation a special laser tube is required. ⋯ For monitoring purposes during jet ventilation or procedures under apnoea, transcutaneous oximetry, ECG recording as well as non-invasive blood pressure measurements at short intervals is mandatory. In addition video monitoring is desirable to allow visual anaesthesiological surveillance of the larynx. The jet ventilator must meet established standards; the option must be available to survey inspiratory peak pressure and end-exspiratory pressure as well as the setting of appropriate alarm limits.