Comparative StudyComparison between the LCx Probe system and the COBAS AMPLICOR system for detection of Chlamydia trachomatis and Neisseria gonorrhoeae infections in patients attending a clinic for treatment of sexually transmitted diseases in Amsterdam, The Netherlands.
Two assays for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae were compared: the LCx Probe system (the LCx system; Abbott Diagnostic Laboratories, North Chicago, Ill.) and the COBAS AMPLICOR C. trachomatis/N. gonorrhoeae system (the COBAS AMPLICOR system; Roche Diagnostic Systems, Branchburg, N. J.). Endocervical swab specimens, male urethral swab specimens, and female and male urine specimens were collected from 503 female and 498 male visitors attending a sexually transmitted diseases clinic in Amsterdam, The Netherlands. ⋯ Sequence analysis of the amplimers obtained by an in-house 16S rRNA PCR of the solely COBAS AMPLICOR system-positive swab specimens revealed neither N. gonorrhoeae nor other Neisseria spp. The COBAS AMPLICOR assay was considered not suitable for screening for infections with N. gonorrhoeae. If this assay is used for detection of N. gonorrhoeae, confirmation of positive results by a reliable test is mandatory.