[Part-body hyperthermia with a radiofrequency multiantenna applicator under online control in a 1.5 T MR-tomograph].
Objective of this study is the integration of a multiantenna applicator for part-body hyperthermia (BSD 2000/3D) in a 1.5 T MR-tomograph (Siemens Magnetom Symphony) in order to perform noninvasive MR monitoring in real time to increase safety and effectiveness of heat treatments. The positioning unit is mechanically coupled to the MR gantry from the back side and the body coil is utilised for imaging. For that purpose, the hyperthermia antenna system (100 MHz, 1.500 W) and the MR receiver (63.9 MHs) have to be decoupled in terms of high frequency (filter) and electromagnetically (emc). ⋯ For 70% of these patients a visualisation of temperature sensitive data during treatment was possible. The evaluated difference images represent a superposition of real temperature -increase and a (temperature-induced) perfusion elevation. The -hybrid approach renders development of part body hyperthermia possible as an MR-controlled intervention in radiology.