International journal of radiation oncology, biology, physics
-
Int. J. Radiat. Oncol. Biol. Phys. · Nov 1999
Optimum timing for image-based dose evaluation of 125I and 103PD prostate seed implants.
Image-based dose evaluation of permanent brachytherapy implants for prostate cancer is important for optimal patient management after implantation. Because of edema caused by the surgical procedure in the implantation, if the dose evaluation is based on the images obtained too early after implantation, dose coverage will usually be underestimated. Conversely, if the images are obtained too late, the dose coverage will be overestimated. This study uses a biomathematical model to simulate edema and its resolution on 29 patients, so that the optimum time to obtain image scans and perform dose evaluation can be investigated and estimated. ⋯ Based on the dynamic model, the optimum timing of image scans for postimplant dose evaluation of prostate seed implantation is 7 weeks postimplantation for 125I implants and about 3 weeks for 103Pd implants. (ABSTRACT TRUNCATED)