Intensive care medicine
-
Intensive care medicine · Feb 1995
Review Comparative StudyCharacteristics of the ventilator pressure- and flow-trigger variables.
Pressure- and flow-triggering are available in the Puritan Bennett 7200ae and Siemens SV 300. Using a mechanical lung model, we described the characteristics of the pressure- and flow-triggered continuous positive airway pressure (CPAP) of both ventilators. In the Puritan Bennett 7200ae, the pressure-triggered CPAP is characterized by the relatively insufficient flow delivery after the triggering, resulting in a greater lung pressure-time product (total PTP) than the flow-triggered CPAP. ⋯ Total PTP is less with pressure- or flow-triggered CPAP of the Siemens SV 300 than that of the Puritan Bennett 7200ae, respectively. The application of small pressure- or flow-triggered pressure support in the Puritan Bennett 7200ae eliminates the difference. The impact of these differences on patient inspiratory muscle work remains to be determined.