Color Doppler velocimetry allows simultaneous imaging by B-scan and Doppler ultrasound. This facilitates estimation of blood velocity in the orbital vessels. A pulse arterial waveform providing blood velocity measurements can be obtained from the optic nerve head. The vascular pattern in this region consists of multiple peripapillary blood vessels and a central artery. The exact contribution of the arteries in this area to velocimetry readings was undetermined. ⋯ The results demonstrate that color Doppler ultrasound of the optic nerve head provides velocimetric measurements of blood flow in the central retinal artery. Although some investigators using different apparatuses have produced velocimetry measurements from the posterior ciliary arteries, in this study reliable measurements were not obtainable form color indices in the peripapillary area.