The International journal of artificial organs
-
This study was conducted to identify the causes of plasma leakage of oxygenators in extra corporeal membrane oxygenation (ECMO). From 1996 through 2002, 91 oxygenators were used in 62 patients undergoing ECMO for respiratory and/or cardiac failure. Several types of oxygenators were used (Medtronic Maxima, Minima, PRF, Medos Hilite). ⋯ The type of oxygenator used has the most overwhelming effect (significantly less leakage with the Medos Hilite). In conclusion, leakage of oxygenators is predominantly caused by the type of oxygenator used. Patient variables (younger age and the number of oxygenators used in one patient) are also significant and allude to an inflammatory process as underlying mechanism of plasma leakage.
-
Comparative Study
Augmentation of abdominal organ perfusion during cardiopulmonary bypass with a novel intra-aortic pulsatile catheter pump.
Current pulsatile pumps for cardiopulmonary bypass (CPB) are far from satisfactory because of the poor pulsatility. This study was undertaken to examine the efficiency of a novel pulsatile catheter pump on pulsatility and its effect on abdominal organ perfusion during CPB. ⋯ Applying the pulsatile catheter pump in the descending aorta is effective in supplying the pulsatile flow to the abdominal organs and results in improved abdominal organ perfusion during the ischemic phase of CPB.