The International journal of artificial organs
-
Randomized Controlled Trial
A significant proportion of patients treated with citrate containing dialysate need additional anticoagulation.
The blood membrane interaction induced during hemodialysis (HD) activates the coagulation system. To prevent clotting and to maintain dialyzer patency, an anticoagulant such as tinzaparin is used. To increase patency of the dialyzers and to reduce the risk of bleeding related to anticoagulation, citrate-containing dialysate has been introduced in Europe. ⋯ A significant proportion of patients treated with citrate-containing dialysate need additional anticoagulation.
-
Comparative Study
Evaluation of alternatives for dysfunctional double lumen central venous catheters using a two-compartmental mathematical model for different solutes.
Double lumen (DL) central venous catheters (CVC) often suffer from thrombosis, fibrin sheet formation, and/or suction towards the vessel wall, resulting in insufficient blood flow during hemodialysis. Reversing the catheter connection often restores blood flows, but will lead to higher recirculation. Single lumen (SL) CVCs have often fewer flow problems, but they inherently have some degree of recirculation. ⋯ The same trend was found for MG and P. However, β2M RR decreased only 1.5% with SL CVC although TSR decrease was 17.2%, while RR decreased 21.1% with DL-low flow although TSR decrease was only 4.9%. In the case of dysfunctional DL CVCs, reversing the catheter connection and restoring the blood flow did not impair TSR, with 10% recirculation. The SL CVC showed suboptimal TSR results that were similar to those of the dysfunctional DL CVC.