Artificial organs
-
The left ventricular assist device (LVAD) with centrifugal pump has two characteristics. One is a pump flow wave of the centrifugal pump, consisting of the pulsatile flow of the native heart and the nonpulsatile flow of the centrifugal pump. The other is that the centrifugal pump fills from the native heart not only in the systolic phase, but also in the diastolic phase. ⋯ Also, it was suggested that LVAD increases the pressure difference between the left atrium and the left ventricle in the diastolic phase. This phenomenon is due to the filling of the left ventricle. In this study it was suggested that as pump assist rate increases, it is more effective to keep cardiac function in the diastolic phase.
-
The rotary blood pump is implanted as a bridge to cardiac transplantation. Mechanical, histological, and biochemical improvements have been described in patients after implantation of left ventricular assist devices (LVADs). Thus, the rotary blood pump might be used as a bridge to recovery of myocardial function. ⋯ This method was studied at normal, hyperdynamic, and heart-failure conditions. There was a high positive correlation between the dp/dt of left ventricular pressure and external stroke work of actuator. This method is useful and simple to estimate cardiac function without pump stop.