Artificial organs
-
Comparative Study
In vitro comparison of support capabilities of intra-aortic balloon pump and Impella 2.5 left percutaneous.
The Impella 2.5 left percutaneous (LP), a relatively new transvalvular assist device, challenges the position of the intra-aortic balloon pump (IABP), which has a long record in supporting patients after myocardial infarction and cardiac surgery. However, while more costly and more demanding in management, the advantages of the Impella 2.5 LP are yet to be established. The aim of this study was to evaluate the benefits of the 40 cc IABP and the Impella 2.5 LP operating at 47,000 rpm in vitro, and compare their circulatory support capabilities in terms of cardiac output, coronary flow, cardiac stroke work, and arterial blood pressure. ⋯ On the other hand, pulsatility was enhanced with the IABP and lowered with the Impella 2.5 LP. The support capabilities of both the IABP and the Impella 2.5 LP strongly depended on the simulated hemodynamic conditions. Maximum hemodynamic benefits were achieved when mechanical circulatory support was applied on a simulated scenario of deep CS.