Artificial organs
-
The aim of this study is to describe the incidence of brain injury (BI) in children with end-stage cardiac failure who were supported with the Berlin Heart EXCOR ventricular assist device (VAD) as a bridge to heart transplantation. Between January 2002 and January 2012, all patients <18 years of age who underwent the implantation of the Berlin Heart EXCOR at Bambino Gesú Children's Hospital were included. A total of 25 patients were included in this study. ⋯ BI is a frequent complication among children supported with EXCOR VAD and is associated with lower weight at implantation. However, our data do not support the association between size and BI. Future prospective multicenter studies are warranted to further help understand the etiology and the impact of BI and to improve functional outcomes for children undergoing EXCOR VAD mechanical support.
-
Cerebral emboli originating in the ascending aorta are a major cause of noncardiac complications following cardiac surgery. The hemodynamics of the aortic cannula has been proven to play a significant role in emboli generation and distribution. The aim of the current study was to perform a thorough numerical investigation in order to examine the effect of the design and orientation of the cannula used during cardiopulmonary bypass on the risk to develop cerebral embolism. ⋯ The TIP2 cannula exhibited a stronger emanating jet, higher shear stress inside the cannula, and highly disturbed flow, which was more stagnant near the clamp region. Current findings support the significant impact of the cannula design and orientation on emboli generation and distribution. Specifically, the straight tip cannula demonstrated a reduced risk of cerebral embolism, which may be pivotal in the clinical setting.