Artificial organs
-
In patients with continuous flow left ventricular assist devices (CF-LVADs) myocardial recovery is uncommon. Given the heterogeneity of the population implanted and low incidence of recovery, the discovery of native left ventricular (LV) recovery and criteria for explantation of CF-LVAD system is not clearly determined. We sought to analyze the characteristics of the patients who underwent CF-LVAD explantation at our institution. ⋯ Clinical evidence of hemolysis and echocardiographic evidence of reduced or absent diastolic flow velocities were common findings in these patients. Over time, patient's native LV function declined in the absence of LVAD (after LVAD explantation). Significant challenges remain in predicting LV recovery and identifying those individuals who have recovered myocardial function significant enough to be explanted.
-
New left ventricular assist devices (LVADs) offer both important advantages and potential hazards. VAD development requires better and expeditious ways to identify these advantages and hazards. We validated in an isolated working heart the hemodynamic performance of an intraventricular LVAD and investigated how its outflow cannula interacted with the aortic valve. ⋯ Aortic pressure influenced central positioning of the outflow cannula in the aortic root. The isolated heart is a simple, accessible evaluation platform unaffected by complex reactions within a whole, living animal. This platform allowed detection and visualization of potential hazards.