Venoarterial extracorporeal membrane oxygenation (VA-ECMO) is indicated in reversible life-threatening circulatory failure with or without respiratory failure. Arterial desaturation in the upper body is frequently seen in patients with peripheral arterial cannulation and severe respiratory failure. The importance of venous cannula positioning was explored in a computer simulation model and a clinical case was described. ⋯ The patient survived, was extubated and showed no signs of hypoxic damage. We conclude that venous drainage from the superior caval vein improves upper body arterial saturation during veno-arterial ECMO as compared with drainage solely from the inferior caval vein in patients with respiratory failure. The results from the simulation model are in agreement with the clinical scenario.
-
The objective of this study was to evaluate an alternative neonatal extracorporeal life support (ECLS) circuit with a RotaFlow centrifugal pump and Better-Bladder (BB) for hemodynamic performance and gaseous microemboli (GME) capture in a simulated neonatal ECLS system. The circuit consisted of a Maquet RotaFlow centrifugal pump, a Quadrox-iD Pediatric diffusion membrane oxygenator, 8 Fr arterial cannula, and 10 Fr venous cannula. A "Y" connector was inserted into the venous line to allow for comparison between BB and no BB. ⋯ This study showed that a RotaFlow centrifugal pump combined with a BB can help to significantly decrease the number of GME in a neonatal ECLS circuit. Even with this optimized alternative circuit, a large percentage of the total hemodynamic energy was lost. The arterial cannula was the main source of resistance in the circuit.
-
Outcomes for extracorporeal membrane oxygenation (ECMO) have been described for patients with single ventricle physiology (SVP) undergoing cavopulmonary connection (Glenn procedure). An alternative surgical pathway for patients with SVP consists of an initial hybrid procedure followed by a comprehensive Stage II procedure. No data exist describing the outcomes of patients requiring ECMO after the comprehensive Stage II procedure. ⋯ Four patients (67%) were discharged alive after ECMO decannulation. Despite being a much more extensive surgical procedure, the morbidity and mortality after ECMO in patients undergoing the comprehensive Stage II procedure are similar to those in patients undergoing the Glenn procedure. If needed, ECMO support is reasonable for patients after the comprehensive Stage II procedure.
-
The objective of this study is to evaluate electrocardiography (ECG)-synchronized pulsatile flow under varying heart rates and different atrial and ventricular arrhythmias in a simulated extracorporeal life support (ECLS) system. The ECLS circuit consisted of an i-cor diagonal pump and console, an iLA membrane ventilator, and an 18 Fr arterial cannula. The circuit was primed with lactated Ringer's solution and packed red blood cells (hematocrit 35%). ⋯ The i-cor console successfully tracked electrocardiographic signals of 12 atrial and ventricular arrhythmias. Our results demonstrated that the i-cor pulsatile ECLS system can be synchronized with a normal heart rate or with various atrial/ventricular arrhythmias. Further in vivo studies are warranted to confirm our findings.
-
The objective of this study was to evaluate the hemodynamic performance and energy transmission of flexible arterial tubing as the arterial line in a simulated pediatric pulsatile extracorporeal life support (ECLS) system. The ECLS circuit consisted of a Medos Deltastream DP3 diagonal pump head, Medos Hilite 2400 LT oxygenator, Biomedicus arterial/venous cannula (10 Fr/14 Fr), 3 feet of polyvinyl chloride (PVC) arterial tubing or latex rubber arterial tubing, primed with lactated Ringer's solution and packed red blood cells (hematocrit 40%). Trials were conducted at flow rates of 300 to 1200 mL/min (300 mL/min increments) under nonpulsatile and pulsatile modes at 36°C using either PVC arterial tubing (PVC group) or latex rubber tubing (Latex group). ⋯ Although total hemodynamic energy (THE) losses were higher under pulsatile mode compared to nonpulsatile mode, more THE was delivered to the pseudopatient, particularly in the Latex group (P < 0.05). The results showed that the flexible arterial tubing retained more hemodynamic energy passing through it under pulsatile mode while mean pressures and pressure drops across the ECLS circuit were similar between PVC and latex rubber arterial tubing. Further studies are warranted to verify our findings.