Artificial organs
-
Rotary ventricular assist devices (VADs) are used to provide mechanical circulatory support. However, their lack of preload sensitivity in constant speed control mode (CSC) may result in ventricular suction or venous congestion. This is particularly true of biventricular support, where the native flow-balancing Starling response of both ventricles is diminished. ⋯ Cardiac output was maintained at sufficient levels by the SLC, with systemic and pulmonary flow rates maintained above 3.14 L/min for all tests. With the CSC, left ventricular suction occurred during reductions in SVR, elevations in PVR, and reduction in circulating fluid simulations. These results demonstrate a need for a physiological control system and provide adequate in vitro validation of the immediate response of a SLC for biventricular support.